PREVAC | Technology for science and industry 13.11.2026 01:24

Gas leak detector GLD10

Categories: ,
Gas leak detector GLD10

Gas Leak Detector GLD10 is a compact leak testing system, equipped with a dedicated electronic device for easy process controlling and monitoring.

The unit is prepared to work in automatic industrial leak detection systems. Allows leak detection at the lowest measurable levels. Thanks to the implementation of the advanced data transmission protocol and easy, user-friendly control interface, GLD10 allows complete integration with the latest production management systems based on the specification of industry 4.0.

Read article about this product

Technical data

GLD10

Min. detectable leak rates (He):

≤ 5E-12 mbar· l/s   for ULTRA MODE (>5 l/s He pumping speed)

≤ 1E-10 mbar· l/s   for FINE MODE (1.7 l/s He pumping speed)

≤ 1E-7   mbar· l/s   for STANDARD MODE

Max. inlet pressure

0.2 mbar  for ULTRA MODE

1.0 mbar  for FINE MODE

20 mbar  for STANDARD MODE

Measuring units mbar· l/s, Pa· m³/s, atm cc/s, Torr· l/s
Response time <1 s
Ion source cathode 2 yttrium/iridium
Vacuum connections DN 16 KF, DN 25 KF
Dimensions 333 x 273 x 270 mm (WxHxD)
Weigth (approx.) 13.5 kg

GLD10-PS

Digital inputs and outputs 10 inputs / 8 outputs
2 analog outputs 0 – 10 V
Communication protocols Modbus RTU – interface: RS232/485
Modbus TCP – interface: Ethernet
User interface 7″ TFT display with touchscreen,
Interface languages English, German, Polish
Dimensions 212.6 x 128.4 x 260.3 mm

(WxHxD), 1/2 of 3HE 19”

Weight (approx.) 3.8 kg

Options

Backing pump TRIVAC D4 B
External calibrated leak range E-7 mbar· l/s
Connecting cables 1 m / 3 m / 5 m / 10 m
Sniffer 4 m

Features

Description

In the spectrometer of the leak detector, the ionized gas atom’s path is bent in a magnetic field, and geometry and control voltage are set for measurement of light gases like Helium 3, 4 AMU or Hydrogen 2 AMU. GLD10 requires a backing pump (available as an option).

 

Applications

Leak tests are performed, for example:

in the industrial vacuum processes:

in the production of automotive parts:

in the production of consumer electronics with heat pumps: