High-power, monochromatized UV source for the study of nanostructures. It use Bragg diffraction to separate narrow band of light wavelengths. During crystal rotation user can find 4 different wavelengths of beam transferred to sample. It happens because each wavelength is reflected by crystal at different angle (Bragg reflection). So called ZERO ORDER beam – creates when monochromator crystal works similar as normal mirror – it reflects multiple wavelengths and gives highest intensitivity of transferred beam. That mode can be used when user does not need monochromated beam, but high intensitivity of UV radiation.
| Mounting flange | DN 40 CF (non-rotatable) |
| Photocurrent | 1 – 12 nA |
| Discharge current | 40 – 300 mA |
| Port aligner | integral , +/- 4° |
| Retractable capillary – distance from focus point to port | 350 mm |
| Capillary I.D. | 1,5 mm (others on request) |
| Insertion length | 345 mm with extended capillary 289 mm with inserted capillary |
| Typical working distance | 5 mm |
| Bakeout temperature | up to 150 °C |
| Working pressure | < 1 mbar |