Based on ellipsoidal quartz crystal and operates according to Bragg Law of X-ray diffraction. Crystal mirror has been installed on specially designed independent retraction, pitch, roll mechanism to precise adjustment working position and two halogen heaters controlled via PID regulator.
The monochromator with a Rowland circle of 500mm diameter for a high X-ray energy resolution has a compact design with differential pumping ports and an optional polymer aluminized window can be installed to prevent from sputtering.
X-ray source based on redesigned dual anode source has been installed on three degrees of motion high precise manipulator. Source has two operation modes – high power up to 600W, and a small spot for high spatial/energy resolution measurement.
High-pressure versions are available:
RMC50 HP5 with a working pressure range up to 5mbar
RMC50 HP50 with a working pressure range up to 50mbar
| Mounting flange | DN 100 CF |
| Crystal area | 200 mm x 100 mm |
| Rowland circle diameter | 500 mm |
| Chamber diameter | 310 mm |
| Chamber port length | 220 mm |
| X-ray source anodes | dual anode Al/Ag (standard) |
| Modes | normal (non-focusing): 1 mm x 4 mm small spot (focusing): 1 mm x 2 mm |
| Voltage | up to 15 kV |
| Power | Al: 200 W (focusing), 450 W (non-focusing) Ag: 300 W (focusing), 600 W (non-focusing) |
| Manipulator X/Y/Z range | ± 6.5 mm / ± 6.5 mm / 25 mm |
| Differential pumping | yes |
| Crystal heating | No |
| Shutter | option |
| Bakeout temperature | up to 150 °C |
| Weight (approx.) | 65 kg |
| Working pressure | UHV -> 5×10-6 mbar HP5: UHV – 5 mbar HP5: UHV – 50 mbar |