Hemispherical energy analyser EA15-HP50

The EA15-HP50 allows high-resolution PES measurements in static and dynamic environments up to 50 mbar.

SKU: 446 Category:

Description

The analyser is equipped with a total number of 11 slits and offers the possibility to choose between best energy resolution and best intensity. According to given photoelectron energy the analyser is set up with up to 9 predefined PE to satisfy customer’s requirements.

Package includes:

▪ EA15-HP50 150 mm hemispherical energy analyser,
RUDI-EA2 high stable and low noise electronics,
SPECTRIUM acquisition and analyser control software for Windows OS computer system.

Technical data

Mounting flange DN 200 CF
Bakeout temperature up to 120 °C
Working distance 0.3-2 mm (dependent on cone opening)
Analyser mean radius 150 mm
Pass energies 5, 10, 20, 50, 100, 200, 250, 300, 500 eV
Energy resolution
XPS
 

< 20 meV FWHM

Kinetic energy range 0.5 – 1500 eV
Acquisition modes fixed, scan
Transmission and angular lens mode:
▫ lens acceptance angle
(transmission mode)
+/- 9°
Maximum energy window in fixed mode 12.5 % of pass energy (for MCP-CCD)
Completely designed of non-magnetic materials

Detectors

•  MCP-CCD detector
– 40 mm diameter dual MCP detector
– 656 energy channels available simultanously
– 494 angular channels available simultanously
– 90 fps
•  7-MCD detector

Additional information

EA15 hemispherical energy analyser can be controlled directly by LabVIEW programming environment.

Note:
The use of a mu-metal analysis vacuum chamber or an inner mu-metal shielded chamber in combination with PREVAC EA15-HP50 hemispherical energy analyser is recommended in order to minimalize influence of external magnetic fields and to maximize performance of the analyser.